search Icon
Current location:HOME>PRODUCT>Biological aerosol detector>Biological aerosol detector
Biological aerosol detector

Biological aerosol detector

【Introduction】Air sampling flow rate: 245L/min, air sampling flow rate: 10-750L/min, detection time: 2-180S

Product model: HM-KWS

  • Product details
  • Detail image
  • Related products
  • Related articles

Product Introduction:

The bioaerosol detector can be used for rapid sampling and detection at the site of biological emergency incidents; It can also be widely used for sampling and testing airborne microorganisms in environments such as disease prevention and control, environmental protection, pharmaceuticals, fermentation industry, food industry, biological cleanliness, and public places. It can also be used for sampling and testing airborne microorganisms in relevant teaching, scientific research, and other institutions.

Product advantages:

1. The bioaerosol detector integrates a high flow acquisition module, a rapid fluorescence detection module, and a cleaning module, achieving fully automatic unmanned detection (can detect at multiple time periods every day), eliminating the need for manual separate sampling and the process of transferring to laboratory testing after sampling is completed;

2. Android system RAM 2G+ROM 16G;

3. High flow air sampling device (dry wall cyclone solid gas separation principle)

4. Adopting MPPT silicon photomultiplier tube detector

5. It can be tested at multiple scheduled times every day;

6. After detecting the automatic message data;

7. Can wirelessly upload data to the cloud platform via WiFi networking;

8. Configure a data management platform that can conduct long-term and short-term evaluation and management analysis;

9. Dual use of AC and DC, convenient for long-term monitoring and flow detection;

10. Optional 4G module and positioning module

Principle Introduction:

The bioaerosol detector is based on the principle of firefly luminescence, using the "luciferase luciferin system" to quickly detect adenosine triphosphate (ATP) in living microorganisms. The ATP content is relatively constant, and the ATP content in the sample can be determined by detecting the luminescence value. Firstly, the microorganisms in the air are enriched in distilled water through a high flow air sampling module, and then a quantitative enzyme solution is added to the liquid. After sufficient reaction, the relative luminescence is tested by the rapid fluorescence detection module, and the corresponding microbial content can be calculated based on the luminescence intensity.

Product parameters:

System: Android 7.1

Air sampling flow rate: 245L/min

Air sampling flow rate: 10-750L/min

Testing time: 2-180 seconds

Detection range: 0-99999999RLUs.

ATP detection limit: 10-10 to 10-12 mol ATP

Temperature range: 18-45 degrees

Collection efficiency: 92%

Standby time:>=10H

Average power consumption: 6W

Power supply: 12V DC power supply

Battery: 12V/10000mAH

Volume: 337 * 163 * 480mm

Weight: 8KG



Article address:https://www.environmentinstrum.com/pro4/21.html