Infrared CO/CO2 Analyzer
Product model: HM-Q3
Product model: HM-Q3
The infrared CO/CO2 analyzer is a high-precision instrument used by our factory to detect toxic and harmful substances in the air of public places. The instrument uses the non dispersive infrared NDIR principle to measure the concentration of carbon monoxide and carbon dioxide in the air, and has the advantages of high measurement accuracy, long service life, and low cross interference. It is an essential instrument in the fields of environmental monitoring and occupational health monitoring.
This instrument is widely used in environmental protection, environmental monitoring, health supervision, occupational health, disease control, and research institutes.
JJG635-2011 Infrared Gas Analyzer for Carbon Monoxide and Carbon Dioxide
HJ965-2018 "Automatic determination of carbon monoxide in ambient air using non diffusive infrared method"
GB/T18204.2-2014 "Methods for hygiene inspection in public places - Part 2: Chemical pollutants"
GB 9801-1988 "Air quality - Determination of carbon monoxide - Non diffusive infrared method"
GBZ/T 300.37-2017 "Determination of Toxic Substances in the Air of Public Places Part 37: Carbon Monoxide and Carbon Dioxide"
1. Adopting Android system, 7-inch touch screen display, with better real-time human-computer interaction.
2. Large storage space, capable of storing over 100000 pieces of data.
3. Equipped with a thermal printer, the current data can be printed after the detection is completed
4. Comes with WiFi, completes detection, can be uploaded to the platform for convenient data monitoring (optional 4G module)
5. Equipped with a USB data interface, data can be imported into a USB drive.
6. Digital flow sensor, real-time data flow display
7. Adopting high-precision infrared sensors, with high measurement accuracy, fast response speed, and short preheating time.
8. Support two methods of nitrogen zeroing and catalytic zeroing.
9. Equipped with internal constant temperature heating function, suitable for outdoor low-temperature environment operation.
10. Built in path switching solenoid valve, automatic switching for zero adjustment and measurement.
11. It has temperature and humidity measurement function, built-in temperature and humidity compensation correction algorithm, and eliminates the influence of temperature changes on measurement data.
12. Display of 0.5 hour mean, 8-hour mean, and daily average; Timed storage function.
1. Detection principle: Non dispersive infrared gas analysis method/Non dispersive infrared method (NDIR)
2. Detecting gases: carbon monoxide (CO), carbon dioxide (CO2) in the air
3. Detection method: Built in pump suction
4. Measurement range: Optional 1, 0~50.0 × 10-6 CO, 0~0.500% CO2 (0-50ppm CO, 0-5000ppm CO2)
Option 2, 0~200.0 × 10-6 CO, 0~1.00% CO2 (0-200ppm CO, 0-10000ppm CO2)
5. Temperature range for use: CO 0-45 ℃, CO2-20 ℃~65 ℃
6. Humidity: 0-95% RH
7. Concentration display automatically switches between ppm, mg/m3, and%
8. Resolution: 0.1 × 10-6 CO; 0.0001% CO2
9. Display screen: 7-inch capacitive touch screen
10. Can store>100000 sets of measurement data
11. Zero point automatic correction technology
12. Linear error: ≤± 2% F · S
13. Repeatability: ≤ 1.0%
14. Span drift: ≤± 2% F · S/3h
15. Zero drift: ≤± 1% F · S/h
16. Carbon monoxide interference: 1250mg/m3CO ≤± 0.3% F · S
17. Response time: CO: t0- t90 ≤ 45S; CO2:≤60S
18. Preheating time: 5 minutes
19. Flow rate: (0.5-2.0) L/min
20. Output interface: USB direct export to USB flash drive
21. Equipped with sound and light alarm functions
22. Power supply: Equipped with a built-in lithium battery, it can be used for both AC and DC, and can be manually switched freely
23. Dimensions: 257.3 * 147 * 300 (mm)
In today's world where air quality is increasingly emphasized, negative oxygen ions, known as &q···...
In modern industry that heavily relies on precision environments, a single speck of dust can shake t···...
In the fields of precision manufacturing and life sciences, air cleanliness is the lifeline of produ···...
Why is precise monitoring of carbon dioxide necessary?Today, with the increasing attention to carbon···...
In highly clean environments such as pharmaceutical workshops, operating rooms, and biological labor···...
In modern industrial production and scientific research, air cleanliness directly impacts product qu···...
In the field of modern industry and technology, air cleanliness is no longer an optional standard, b···...
The airborne dust and bacteria sampler is an efficient microbial detection device designed based on ···...